            | Hematology Analyzer |  
            |  |  
            | XFA 6100 Animal Auto Hematology Analyzer |  
            | Used to detect the parameters of the series of red blood cells and   hemoglobin,platelet and leukocyte in the blood, it is an in-vitro diagnostic   equipment which is applied to quantitative analysis of blood cells of clinical   and laboratory. |  |  
            |  |  
            | 
              
                | Test Items | WBC, LY#, MO#, GR#, LY%, MO%, GR%,    HGB, RBC, HCT, MCV, MCH, MCHC, RDW—CV, RDW—SD, PLT, PCT, MPV, PDW, WBC    Histogram, RBC Histogram, PLT Histogram |  
                | Background Count | WBC: ≤0.3×109/L      RBC: ≤0.03×1012/L   HGB: ≤1g/L   PLT:    ≤10×109/L |  
                | Linearity | Test    Item | Test    Range | Allow    Difference of Linearity |  
                | WBC | 0~6.0×109/L | ±0.3×109/L |  
                | 6.0~99.9×109/L | ±5% |  
                | RBC | 0~0.99×1012/L | ±0.05×1012/    L |  
                | 1.0~9.99×1012/L | ±5% |  
                | HGB | 0~99 g/L | ±2g/L |  
                | 100~300 g/L | ±2% |  
                | PLT | 0~99×109/L | ±10×109/L |  
                | 100~999×109/L | ±10% |  
                | Repeatability | Test    Item | CV(%) | Teat    Range |  
                | WBC | ≤2.0 | 7.0~15.0×109/L |  
                | RBC | ≤0.5 | 3.5~6.0×1012/L |  
                | HGB | ≤1.5 | 110~180g/L |  
                | PLT | ≤4.0 | 150~500×109/L |  
                | MCV | ≤0.5 | 80~110fL |  
                | Sample Volume | Twig blood: 20μl |  
                | Dilution Ratio | WBC/HGB: 1:500   RBC/PLT:    1:50000 |  
                | Interfaces | RS232 joint, USB, PS2, Parallel    port |  
                | Test Speed | 60 samples/hours |  
                | Continuous Work Time | >8h |  |  
            |  |  
            |  |  
            | XF9030B Semi-automatic Hematology Analyzer |  
            |  |  
            | XFA6100 is used to detect the parameters of the series of red blood cells and   hemoglobin,platelet and leukocyte in the blood ,it is an in-vitro diagnostic   equipment which is applied to quantitative analysis of blood cells of clinical   and laboratory. |  |  
            |  |  
            | 
              
                | Test Items | RBC, HGB, WBC, LY#, MO#,
                  GR#, LY%, MO%, GR%, HCT, MCV, MCH,
                  MCHC, RDW, PLT, MPV, PCT, PDW 
 |  
                | Linearity Range | Test    Item | Test    Range |  
                | RBC | 1.00~9.99×10 ¹² /L |  
                | HGB | 20g/L~200g/L |  
                | WBC | 0.50~99.0×10 9/L |  
                | PLT | 50~999×10 9/L |  
                | Precision (CV) | Test    Item | CV(%) |  
                | RBC | ≤2.5% |  
                | HGB | ≤1.5% |  
                | WBC | ≤3.5% |  
                | LY% | ≤6% |  
                | MO% | ≤12% |  
                | GR% | ≤6% |  
                | MCV | ≤2% |  
                | PLT | ≤5% |  
                | MPV | ≤3% |  
                | Dilution Ration | WBC / HGB | 1:500 |  
                |  | RBC / PLT | 1:50000 |  
                | Sample Volume | Periphery/whole blood: 20μl |  
                | Packing details | 430mm×410mm×430mm |  
                | Weight | N.W.11kgs   G.W.16kgs |  
                | Reagents | HB Non-cyanogen; Hemolysin 2×15ml |  
                |  | HB Diluent 4L |  
                |  | HB Washing Solution 4L |  |  
            |  |  
            |  |  
            |  |  
            |  |  
            |  |  
            |  |  |